Why Pulse Oximeters Fail to Detect Carbon Monoxide Poisoning

Standard pulse oximeters use a two-wavelength optical system that cannot distinguish between oxygen and carbon monoxide, causing the device to report a falsely normal oxygen saturation even when a patient is severely hypoxic.

By Hui Lin

Share this story

Listen

In short

  1. Standard pulse oximeters use only two wavelengths of light (660 nm and 940 nm) to estimate oxygen saturation, making them blind to other molecules.
  2. Carbon monoxide binds to hemoglobin to form carboxyhemoglobin, which absorbs 660-nanometer red light almost identically to normal oxygenated blood.
  3. Because the device cannot distinguish between the two molecules, a patient dying of carbon monoxide poisoning will display a falsely normal SpO2 reading of 98 to 100 percent.

On July 6, 2026, emergency medicine physician Dr. Erik Axene published a stark warning to prehospital care providers regarding winter emergency calls. He described a scenario where a family complains of severe headaches and confusion, yet their vital signs appear perfectly stable.

"Everyone's SpO2 reads 98 to 100 percent," Axene wrote, detailing the clinical presentation of carbon monoxide exposure. "This is carbon monoxide poisoning until proven otherwise, and your pulse oximeter is the least useful tool you have right now."

Carbon monoxide is notoriously difficult to detect in the field because it is colorless, odorless, and mimics dozens of other conditions. The most dangerous aspect of the toxidrome is its ability to completely blind the primary respiratory monitoring tool used by medical professionals.

A patient actively dying from cellular asphyxiation will consistently display a normal or near-normal oxygen saturation reading on a standard hospital monitor. The physiological mechanism behind this failure lies in the fundamental physics of how light interacts with human blood.[4]

How standard pulse oximeters see light

Since the American Society of Anesthesiologists adopted pulse oximetry as a minimum monitoring standard in 1986, the technology has relied on a simple optical principle. The device aims to provide a continuous, noninvasive assessment of functional arterial oxygen saturation.

A standard pulse oximeter utilizes an electronic processor and a pair of small light-emitting diodes facing a photodiode. These components are typically placed across a translucent part of the patient's body, such as a fingertip or an earlobe.[3]

The device emits light at two specific frequencies: a red wavelength at 660 nanometers and an infrared wavelength at 940 nanometers. Absorption of light at these two distinct wavelengths differs significantly between blood loaded with oxygen and blood lacking oxygen.[3]

Standard pulse oximeters rely on just two wavelengths of light to estimate oxygen saturation.

Oxygenated hemoglobin absorbs more infrared light at 940 nanometers and allows more red light to pass through. Conversely, deoxygenated hemoglobin allows more infrared light to pass through while absorbing a greater amount of the 660-nanometer red light.[1][3]

The device relies on the Beer-Lambert law, which states that the absorption of light passing through a substance is directly proportional to the concentration of the attenuating species. During pulsatile blood flow, the expansion of blood vessels alters the hemoglobin concentration.[1]

By subtracting the non-pulsatile elements of tissues and venous blood, the microprocessor isolates the arterial component. It then calculates a modulation ratio, known as R, by dividing the pulsatile red light absorbance by the pulsatile infrared light absorbance.[1]

This calculated R ratio is then compared against a standardized calibration curve stored in the device's database. The empirical algorithm translates the optical ratio into the SpO2 percentage displayed on the clinical monitor, assuming only two hemoglobin states exist.[1]

The 660-nanometer blind spot

The fatal flaw in this two-wavelength system emerges when a third hemoglobin species enters the bloodstream. Normal human blood contains four hemoglobin species, but standard pulse oximeters are physiologically incapable of distinguishing dyshemoglobins from normal oxygen-carrying molecules.

Carbon monoxide binds to the iron atoms in hemoglobin with an affinity approximately 200 to 250 times stronger than that of oxygen. When inhaled, the toxic gas rapidly displaces oxygen, forming a stable compound known as carboxyhemoglobin.[3]

Carboxyhemoglobin possesses a unique light absorption profile that confounds the standard two-wavelength measurement. At the 660-nanometer red wavelength, the light absorbance curve of carboxyhemoglobin overlaps almost perfectly with the absorbance curve of normal oxyhemoglobin.[3]

Carboxyhemoglobin absorbs 660-nanometer red light almost identically to normal oxygenated blood, blinding the sensor.

"The pulse oximeter will see the reflected light for carboxyhemoglobin and oxyhemoglobin similarly when measuring in the red light spectrum," notes a technical review by Storm Anesthesia. The device measures a combined reflection, entirely unable to separate the two molecules.

Furthermore, carboxyhemoglobin absorbs virtually no light in the 940-nanometer infrared spectrum. Because it does not contribute to the infrared absorbance, the denominator of the R ratio remains unchanged, leaving the device's microprocessor completely unaware of the toxic substitution.

Calculating the false normal

When the pulse oximeter computes the ratio between the 660-nanometer and 940-nanometer signals, the identical red absorbance and absent infrared absorbance of carboxyhemoglobin mimic the exact optical signature of pure oxyhemoglobin. The algorithm processes the data exactly as programmed.[3][4]

The resulting mathematical output forces the device to report a falsely elevated SpO2 reading. The monitor displays a number that reflects the total percentage of hemoglobin bound to any molecule, rather than the percentage bound specifically to oxygen.[3]

The magnitude of this error scales directly with the severity of the poisoning. A patient whose bloodstream contains 40 percent carboxyhemoglobin and only 60 percent oxyhemoglobin will routinely display an SpO2 reading of 98 to 99 percent.

"Their hemoglobin is bound to a molecule that will not offload to tissues," Axene explained. The monitor reassures the clinician that the blood is fully saturated, masking the reality that nearly half of the patient's oxygen-carrying capacity has been destroyed.

This optical illusion means a standard pulse oximeter will never detect carbon monoxide poisoning, regardless of how severe the hypoxia becomes. The patient will suffocate at the cellular level while the digital display insists their respiratory status is perfect.[4]

Because the device cannot distinguish between the two molecules, a lethal carbon monoxide level registers as a healthy oxygen level.

Hypoxia hidden in plain sight

The diagnostic challenge extends beyond noninvasive fingertip monitors. Standard arterial blood gas analysis is equally inadequate for diagnosing carbon monoxide toxicity, as the oxygen saturation reported in these tests represents dissolved oxygen in the plasma.[2]

Dissolved oxygen levels remain entirely unaffected by the concentration of carboxyhemoglobin inside the red blood cells. According to the Merck Manuals, relying on arterial blood gases or standard pulse oximetry will consistently fail to identify the life-threatening deficit.[2]

A normal hemoglobin count of 15 grams per deciliter can typically bind about 20 milliliters of oxygen per deciliter of blood. When carbon monoxide occupies those binding sites, the total oxygen content plummets, but the dissolved plasma oxygen remains normal.[3]

This physiological reality makes clinical suspicion the most vital diagnostic tool for first responders. Recognizing the environmental clues and the classic symptoms of headache, nausea, and confusion must override the falsely reassuring numbers on the vital sign monitor.

Measuring beyond two wavelengths

To accurately detect carboxyhemoglobin noninvasively, medical technology had to move beyond the two-wavelength limitation. Modern pulse co-oximeters utilize eight or more distinct wavelengths of light to differentiate between multiple hemoglobin species simultaneously.

By measuring light absorbance across a wider spectrum, these advanced devices can isolate the unique optical signatures of carboxyhemoglobin and methemoglobin. The microprocessor then calculates the specific concentration of each species, displaying the carbon monoxide level as an SpCO percentage.

Devices like the Rad-57 signal extraction pulse co-oximeter have been approved for clinical use to assess these specific percentages. However, studies published in Frontiers in Medicine suggest that even advanced noninvasive SpCO screening has limitations in accuracy.

Illustration: Advanced pulse co-oximeters use eight or more distinct wavelengths of light to differentiate between multiple hemoglobin species.

The sensitivity of noninvasive co-oximetry is sometimes insufficient to serve as a definitive screening test in emergency departments. False positives and false negatives can still occur, complicating the triage process for patients with suspected exposure.

The ultimate diagnostic standard remains laboratory blood co-oximetry. This invasive test spectrophotometrically measures light transmission through a drawn venous or arterial blood sample at multiple discrete wavelengths, providing a highly accurate measurement of true oxygen content.[2]

Until multi-wavelength technology becomes universal in prehospital care, standard pulse oximeters will continue to operate with a fatal blind spot. Clinicians must remember that a 660-nanometer light beam cannot tell the difference between a breath of fresh air and a lethal dose of exhaust.[4]

How we did this

Method
Deriving the mathematical failure of the standard pulse oximetry algorithm by comparing the light absorption profiles of carboxyhemoglobin against the device's expected R-ratio inputs at 660 nm and 940 nm.
What we found
The standard R-ratio calculation (Pulsatile 660 / Pulsatile 940) fails because carboxyhemoglobin inflates the 660 nm numerator while leaving the 940 nm denominator unchanged. This mathematically forces the microprocessor to interpret the toxic compound as pure oxyhemoglobin, outputting a false 99% saturation even when the patient is severely hypoxic.
What we worked from
  • Red light (660 nm) absorption profile: Identical for HbO2 and HbCO
  • Infrared light (940 nm) absorption profile: Zero for HbCO
  • Clinical SpO2 output at 40% HbCO: 99%
Limits of this analysis
This mathematical derivation applies strictly to standard two-wavelength pulse oximeters and does not account for the proprietary signal extraction algorithms used by modern multi-wavelength co-oximeters.

Key terms

Pulse Oximeter
A noninvasive medical device that estimates the oxygen saturation of blood by shining red and infrared light through the skin.
Carboxyhemoglobin (COHb)
A stable complex formed when carbon monoxide binds to hemoglobin in red blood cells, preventing them from carrying oxygen.
SpO2
The estimated percentage of oxygenated hemoglobin in the blood, as measured by a standard pulse oximeter.
Co-oximetry
An advanced measurement technique that uses multiple wavelengths of light to distinguish between different types of hemoglobin, including those bound to carbon monoxide.
Beer-Lambert Law
A physical law stating that the absorption of light passing through a substance is proportional to the concentration of the molecules absorbing it.

Frequently asked

Can I use a smartwatch to detect carbon monoxide poisoning?

No. Smartwatches and consumer fitness trackers use the same basic two-wavelength optical technology as standard medical pulse oximeters, meaning they are equally blind to carboxyhemoglobin and will report a falsely normal oxygen level.

Why doesn't carbon monoxide affect the infrared light reading?

The molecular structure of carboxyhemoglobin simply does not absorb light at the 940-nanometer wavelength, making it virtually invisible to the infrared sensor used in standard oximetry.

How long does carboxyhemoglobin stay in the blood?

The half-life of carboxyhemoglobin is about four to five hours when breathing normal room air, but this drops to around 40 to 80 minutes when a patient is treated with 100 percent high-flow oxygen.

Do all hospitals have multi-wavelength co-oximeters?

While most hospital laboratories have benchtop co-oximeters for analyzing drawn blood samples, noninvasive fingertip co-oximeters are still not universally carried by all ambulance crews or standard triage desks.

Viewpoints in depth

Emergency Medical Responders

Focus on the clinical danger of false vital signs in the field and the need to rely on environmental clues over digital monitors.

For paramedics and emergency physicians, the pulse oximeter's blind spot represents a critical vulnerability in prehospital care. Responders are trained to trust their monitors to detect hypoxia, but carbon monoxide poisoning actively weaponizes that trust. Medical educators like Dr. Erik Axene stress that when a patient presents with headache, nausea, or confusion in a high-risk environment—such as a closed house in winter—a normal SpO2 reading must be entirely disregarded. The clinical presentation and environmental context are the only reliable indicators until advanced diagnostics are available.

Medical Technologists

Emphasize the physical limitations of two-wavelength spectrophotometry and the mathematical algorithms that drive standard oximeters.

Engineers and technologists view the false normal reading not as a malfunction, but as a rigid adherence to the device's programmed physics. Standard oximeters operate on an empirical algorithm built around the Beer-Lambert law, assuming only oxyhemoglobin and deoxyhemoglobin exist in the blood. Because carboxyhemoglobin mimics the 660-nanometer absorbance of oxygenated blood while remaining invisible at 940 nanometers, the microprocessor calculates a modulation ratio that perfectly matches a 99 percent saturation curve. The device is functioning exactly as designed; it simply lacks the optical vocabulary to recognize the poison.

Clinical Diagnosticians

Advocate for multi-wavelength co-oximetry and invasive arterial blood gas analysis to accurately measure true oxygen content.

Laboratory specialists and diagnosticians emphasize that true oxygen content can only be verified by expanding the spectrum of measurement. They advocate for the broader deployment of noninvasive pulse co-oximeters, which use eight or more wavelengths to isolate the specific optical signatures of dyshemoglobins. However, because even advanced noninvasive screening can yield false positives, diagnosticians maintain that invasive laboratory blood co-oximetry remains the gold standard. They caution against relying on standard arterial blood gas tests, which only measure dissolved plasma oxygen and fail to detect the hemoglobin deficit caused by carbon monoxide.

Emergency Medical Responders 40%Medical Technologists 30%Clinical Diagnosticians 30%
Emergency Medical Responders
Focus on the clinical danger of false vital signs in the field and the need to rely on environmental clues over digital monitors.
Medical Technologists
Emphasize the physical limitations of two-wavelength spectrophotometry and the mathematical algorithms that drive standard oximeters.
Clinical Diagnosticians
Advocate for multi-wavelength co-oximetry and invasive arterial blood gas analysis to accurately measure true oxygen content.

Perspectives this story doesn't cover

  • Pulse Oximeter Manufacturers
  • Patients with Chronic CO Exposure

Sources

Source coverage

4 outlets

3 viewpoints surfaced

Emergency Medical Responders 40%Medical Technologists 30%Clinical Diagnosticians 30%
  1. [1]Life in the Fast LaneMedical Technologists

    Pulse Oximetry

    Read on Life in the Fast Lane →
  2. [2]Merck ManualsClinical Diagnosticians

    Carbon Monoxide Poisoning

    Read on Merck Manuals →
  3. [3]University of IowaMedical Technologists

    Pulse Oximetry: Basic Principles and Interpretation

    Read on University of Iowa →
  4. [4]Factlen Editorial Team

    Synthesis by Factlen editorial team

    Read on Factlen Editorial Team →
Listen

Comments

Stay informed

Every angle. Every day.

Get Guides stories with full source coverage and perspective breakdowns, free every day.